Computer Compatible Manometer Bank

Categories:

Electronic multiple channel pressure measurement system designed to replace liquid manometer banks. 16 simultaneous measurements displayed on a PC (not supplied). Sophisticated software included,...


title icon
Product

Description

  • Electronic multiple channel pressure measurement system designed to replace liquid manometer banks.
  • 16 simultaneous measurements displayed on a PC (not supplied).
  • Sophisticated software included, for logging, recording and displaying data.
  • 16 channels of 350mm H20 capacity, or 14 channels of 350mm plus 2 channels of 125mm H20.
 Sophisticated software is provided with facilities to:
  • Display the data from the 16 channels in real time
  • Record data either on operator request, or at constant sampling periods
  • Define the data sampling periods to be used
  • Record data onto the computer for lateral retrieval, including the option to record the data in Excel format
  • Display tables and graphs of the captured data, withreal time updates
  • Calibrate and scale the individual measurement channels > Zero the data channels before taking measurements
Specification:
General Purpose Channels:
  • Range: 0 to 350mm H20
  • Repeatability: 0.5mm H20
  • Absolute accuracy: 0.5% FSD
  • Media: Wet or dry, non-corrosive
High Sensitivity Channels:
  • Range: 0 to 125mm H20
  • Repeatability: 0.1mm H20
  • Absolute accuracy: <0.5% FSD
  • Media: Dry air only (all figures typical)
 


title icon
Product

Reviews

add Review
reviews

    No Review Yet.

quick overview :

  • Electronic multiple channel pressure measurement system designed to replace liquid manometer banks.
  • 16 simultaneous measurements displayed on a PC (not supplied).
  • Sophisticated software included, for logging, recording and displaying data.
  • 16 channels of 350mm H20 capacity, or 14 channels of 350mm plus 2 channels of 125mm H20.
 Sophisticated software is provided with facilities to:
  • Display the data from the 16 channels in real time
  • Record data either on operator request, or at constant sampling periods
  • Define the data sampling periods to be used
  • Record data onto the computer for lateral retrieval, including the option to record the data in Excel format
  • Display tables and graphs of the captured data, withreal time updates
  • Calibrate and scale the individual measurement channels > Zero the data channels before taking measurements
Specification:
General Purpose Channels:
  • Range: 0 to 350mm H20
  • Repeatability: 0.5mm H20
  • Absolute accuracy: 0.5% FSD
  • Media: Wet or dry, non-corrosive
High Sensitivity Channels:
  • Range: 0 to 125mm H20
  • Repeatability: 0.1mm H20
  • Absolute accuracy: <0.5% FSD
  • Media: Dry air only (all figures typical)
 